Parallel Chiral Column Screening
In chiral column screening, it is essential that the conditions for
separating chiral substances are found quickly and effectively. As
a result, many large pharmaceutical companies have recently decided
to equip their chiral column screening departments with an
8x Sepmatix system. The 8x Sepmatix system for chiral column screening can separate
a substance on eight different chiral separation columns
simultaneously. This significantly increases sample throughput.
The system consists of three modules: an 8x FlowControl low-viscosity
version with flow rates of between 0.25 to 1.50 ml/min; an
autosampler with syringe pump which injects samples from standard
microtiter plates or deep-well microtiter plates onto eight columns
(also suitable for vial racks); and an 8x diode array detector
with eight analytical flow cells applicable for the spectral range of
200 to 600 nanometers. The required space for the 8x Sepmatix system for chiral column
screening is only 55 x 55 x 90 centimeters (W x d x h). The 8x Sepmatix system for chiral column screening can be fitted
with the Sepmatix EluentRack, which is designed for a maximum
of four solvent selection valves (each with six positions) and up
to 24 solvent bottles (1 liter) or 18 fl asks (2.5 liters). the required
space for the sepmatix eluentrack is 55 x 55 x 90 centimeters
(W x d x h).
The parallel screening system shown above
is assembled from the following modules:
- HPLC Pump,Knauer
- Solvent Selection Rack- 24 solvents, Sepiatec (right)
- Sepmatix 8x FlowControl, Sepiatec (left)
- Sepmatix 8x Autosampler, Sepiatec (left)
- Sepmatix 8x DAD Detector, Sepiatec (left)
- Sepmatix 8x Column Oven, 10-70 ° C, Sepiatec (left)
HPLC Conditions:
- Flowrate: 1 ml/min per channel
- Eluents: Hexane with different additives Ethanol, Methanol
and Isopropanol
- Detection: DAD 200-600 nm, online 210 nm
- Injection: 5μl in each channel
- Columns: 8x 250x4.6 mm
Settings for the Solvent Selection Rack
The solvent rack providing a selection
of six solvents at each of the four low
pressure gradient input valves of the
pump gives the option to connect 24
different solvents to the system.
To ease the process of the method
programming first a table of all
solvents is entered in the software.
After setting the right solvent for each
inlet, only the percentage per low
pressure gradient inlet valve has to
be set in the method.
Exchange of solvent can be done in less than one minute due to the fact that
only the solvent in the short connections from the solvent selection valve to
the pump has to be exchanged using the flowrate for eight channels.
Running coated and immobilized polysaccharide CSPs in the same system
Since coated CSPs would avoid the
use of some solvents, the Sepmatix
system allows switching off the
channels with these columns for the
runs with solvents that might damage
the coated phases.
A valve integrated in the FlowControl
unit completely blocks the flow for
these channels and no damaging
solvent reaches the coated columns.
Valves for channels disabled in the Sample Table are colored yellow in the Online screen.
Security System for coated polysaccharide
To avoid damaging of coated polysaccharide
columns the Sepmatix
Control Software has a security
system that allows defining damaging
solvents for each column. With this
information the software ensures that
no damaging solvent can be used with
a wrong column. If wrong solvents are chosen by the
user, the software will not start the run
to avoid any damages to the
columns.
Purge setup to flush out traces of "damaging" solvents
To ensure that no traces of damaging solvents that might still be trapped in
the FlowControl can reach the columns, an automatic purge is integrated to
flush all the solvent out of the FlowControl.
The necessity to flush is determined by the software without user
interference.
Sepmatix 8xColumnSelector
To combine an interesting new idea like parallel screening with an
established approach might even enhance the procedure further. In this
case a set of column selectors is added to the parallel HPLC system.
With this setup in each of the eight channels six different columns can be
installed. This setup allows 48 different columns to be screened
automatically without user intervention.
This 48 columns can either have different CSPs or the same columns can be
used in neutral, acid and basic mode.
Results
With the Chiral Column Screening Wizard Software it is possible to
reprocess up to 80 chromatograms (10 sepmatix runs) in a few minutes.
Visual inspection of the chromatograms shows the best separation on
Column 4 with the solvents used in Method 13. The resulting parameters
from the integration are used to build up a hit list of separations.
Conclusions
With the upcoming of more stationary phases for the separation of chiral
samples, enhancing the throughput of column screening is a key issue for a
thorough test of possible combinations. With a parallel HPLC system the
process can be speeded up. Thus the influence of other parameters and
further solvents can also be investigated.
|